Biological Safety Cabinet

Biological Safety Cabinets [“Class II”] an integral factory assembled system, permitting
total disinfections. These are designed to use simultaneous primary biological and
chemical containment. The re-circulation of cabinet air through a second HEPA Filter
(High Efficiency Particular Air-Mini Pleat type) otters a highest level of cleanliness. Air
is drawn through pre-filter and is made to pass through HEPA Filter having efficiency to
pertaining all airborne particles of size 0.3 microne and larger.

Using a dynamic balanced motor with blower assembly system [1/5HP-1440 RPM]
operates with minimum noise level and the exhaust duct [1/4 HP motor-2880 RPM] Idled
On the back bottom 01 the unit. The basic unit made of Mild Steel powder Coating which
includes standard accessories like static pressure manometer.
DownLoad Brochure

Description

HEPA Filters for cab net [Mini pleat type] Pre-filters. one UV light for the work area. one
coach for gas. It is suitable for handling medium to low risk organisms encountered in
Research. Investigation. Labs in all branches of Medical. Agriculture. and Pharmaceutical
Sciences. The equipment is built with an exhaust system on the top of the unit having 4’
dia PVC duct for the provision of connecting the pipe. in addition to normal features of
Vertical Laminar Flow class 1 it is provided with following unique features.

  • Front door provided with UV stabilized polycarbonate sheet in foldable
  • FIltered Exhaust: Protests ambient environment, avoids build up. heat. for
  • Supplementary Air Barrier:  Prevents of contamination through make open.
  • Built In U.V germicidal light:  facilitates sterilization of working area before and after
    use , U.V Tube (Philips).
  • Cleanliness “CLASS ll“ with an efficiency of 99.99%
Model No Size of worktableSS Top Outer Dimension
EBSC-1 2″ X 2″ 2″ 29″ X 36″ 76″
EBSC-2 3″ X 2″ 2″ 41″ X 36″ 76″
EBSC-3 4″ X 2″ 2″ 53″ X 36″ 76″
EBSC-4 5″ X 2″ 2″ 65″ X 36″ 76″
EBSC-5 6″ X 2″ 2″ 77″ X 36″ 76″

NB:  In accordance to its safety performance necessities and stipulations. the Sc Safety
Cabinet (Class II) can be constructed in the following types
Type A: 70% air (approx) re circulating & 30% exhausting
Type B1: 30% air (approx) re circulating & 70% exhausting
Type B2: 100% exhausting and no re circulation
Bio Safety cabinet (Class I) can be supplied

OPTIONAL :

  • Virus burn out chamber
  • Dwyer magnehelic gauge
  • Glass Bead Sterilizer 2 ̎dia 4 ̎ depth
  • Infra-red activated sterilizer can be supplied
  • For Type B-2, Carbon HEPA Filter (for filtering air prior to discharge the outside area)